Computer Software Assurance: New Way of Computerized System Validation

Updated: Mar 26, 2022

Now a days everyone is shifting from CSV to CSA, a risk based and more efficient validation approach as compared to generating piles of documents which is un-necessary, only to satisfy regulatory bodies. However regulatory bodies does not want to generate piles of documents for validating computerized system rather than they want companies to have defect free solution to improve their Product Quality, Patient Safety and Data Integrity.


During a 2011 review of medical device quality data, the FDA’s Center for Devices and Radiological Health (CDRH) noticed a variety of widespread manufacturing risks that were impacting product quality. A few of these risks included:

  • An industry focus on regulatory compliance as opposed to adopting best quality practices

  • Lack of adoption of automation and digital technologies, with manufacturers choosing instead to continue running long-outdated versions of software

  • Virtually no competitive market around medical device quality


To address these issues, the CDRH, in collaboration with the Center for Biologics Evaluation and Research (CBER) and the Center for Drug Evaluation and Research (CDER), is planning to release a new guidance document later this year (2021) entitled Computer Software Assurance for Manufacturing and Quality System Software. This guidance will create opportunities for streamlining documentation by shifting the focus of CSV processes towards critical thinking, risk management, patient and product safety, data integrity, and quality assurance.

Even though this guidance is being developed for the medical device industry manufacturing, the FDA has indicated that it will be suitable for R&D, clinical, and other groups within pharmaceutical, biotech and medical device companies that are currently following electronic records and signatures and CSV regulations.



Difference between CSA and CSV

Computer software validation (CSV) as defined by the FDA is a documented process to demonstrate that computer software works according to its intended design. Computer software assurance (CSA) is a risk-based approach that limits testing to functions directly impacting product quality and patient safety. CSA also encourages companies to use the software vendor’s documentation to reduce the testing burden and deploy applications faster.

According to the FDA, software not used in a product (e.g., off-the-shelf production manufacturing software) is not subject to the same rigorous requirements of computer systems validation that have been used in process validation. Instead, manufacturers can use computer systems assurance to target verification and validation testing for only high-risk activities.

The FDA’s Center for Devices and Radiological Health (CDRH) has prioritized guidance on CSA for QMS software for publication in 2021 or 2022. It will provide specific guidelines to define CSA and provide direction for making this transition.

Challenge of Computerized System Validation

Encouraging investment in technology is a priority for the FDA as it acknowledges digital transformation as a catalyst for reducing quality issues and patient risk. Conversely, the agency’s Case for Quality initiative to identify the practices of high performers in the medical device industry, identified software validation as a leading roadblock to technology adoption.

Unfortunately, the industry misinterpreted 21 CFR Part 820.75 process validation and applied those requirements to software used in a life science company. Manufacturers in turn subjected all software used in their processes—from software as a medical device (SaMD) to document management software—to extensive validation testing.

The unintended consequence was steep computer software validation implementation and change management costs. The prohibitive costs often meant life sciences companies avoided technology investments altogether—the exact opposite of what the FDA is trying to promote.

What is CSA

The most obvious question is: What is CSA, and how does it differ from traditional CSV?

The current process, CSV, focuses on producing accurate, approved documentation to present to auditors, then testing, then assurance needs, and finally critical thinking.

The CSA methodology is the opposite; It emphasizes critical thinking (risk-based), assurance needs, testing activities, and documentation, in that order. In a nutshell, the goal of CSA is to "right-size" validation activities, placing the focus on what directly impacts patients or product quality.

What’s the FDA looking for?

Although there are subtle differences between CSV and CSA, such as ad hoc testing, CSA still promotes a risk-based, least burdensome approach consistent with the 2002 guidance. In reality, the FDA isn’t asking us to do something that’s incredibly different from what we’re doing now. They’re just asking us to do it better by putting critical thinking at the forefront and leveraging the powerful technology tools we have at our disposal.

What is critical thinking?

Critical thinking is a complex subject that’s difficult to define. Generally speaking, it is the rational, skeptical and unbiased analysis of factual evidence. The ability to reason is a critical thinking skill. Reasoning occurs when we use our knowledge of one thing, process or statement to determine if another thing, process or statement is true.

Logic is another key element of critical thinking. People often confuse reason with logic, but they’re not the same. Reason is subject to personal opinion or bias; logic is strictly fact-based. Logic states if A=B and B=C, then A=C. It doesn’t matter how you feel about A, B or C; it’s a fact. We call this factual statement a premise. Premises help us draw conclusions.

Logical thinking is about connecting one thing with another. Critical thinking is about making sure those connections are valid and not influenced by assumptions or bias. You must become a logical thinker before you can become a critical thinker. But how?

Computer Software Assurance : A Ris Based Approach

CSA involves applying different degrees of validation testing according to the risk associated with the software. CSA can be broken down into four broad steps:

1. Identify the intended software use. Does the software impact product quality or patient safety? If not, you don’t need the same level of assurance as with software that impacts the product or patient safety. For example, software such as document management, change management, and audit management doesn’t require the same level of testing as software within a device.

2. Determine areas or functions that can impact product quality, patient safety, or system integrity. Computer systems validation focuses on calculating the severity of impact and probability of failure, leading to a prioritization of risk. The recommended process with computer systems assurance, however, is calculating the risk/impact on patient safety and product quality, as well as the implementation method of the software functionality.

3. Leverage vendor documentation where possible. Companies shouldn’t waste time reproducing audited vendor documentation and validation. If the vendor has quality documentation and validation in place, use it.

4. Conduct scripted and unscripted testing activities as needed based on risk. Using a risk-based approach requires experts to understand business processes and evaluate the system’s functionality to get more value from their software.


Figure 1: Sample risk-based impact score based on the type of implementation method.


Figure 2: Computer software implementation definitions.



Figure 3: Validation assurance activities and related risk ratings.


Unscripted testing is used on lower-risk systems that do not directly impact the product or patient safety. Detailed test scripts aren’t needed, but companies should document a test objective and pass/fail results.

Scripted testing is for higher-risk systems that directly impact the product or patient safety. The traditional step-by-step procedure with expected results and pass/fail should be documented.

What the Transition to Computer Systems Assurance Means for the QMS

In terms of QMS software, which is not part of the product, the risk to product quality and patient safety is far lower and therefore does not require the same level of testing. GAMP 5 guidance acknowledges the evolution of industry software, recognizing distinct levels of risk associated with different types of software. The most popular industry software category is Category 4 configured software, which allows companies to use a set of verified tools to make software changes without coding to meet business process needs and requirements.

In other words, the shift to CSA saves time, money, and resources. You don’t need to document and perform full validation for out-of-the-box software that’s already been done by the software vendor.

Here’s a practical example: While manufacturers should test critical elements such as user access roles, that does not necessarily require generating a 100-page test script. Instead, companies can leverage the supplier’s testing and verify the functionality in an end-to-end use case test scenario. In this situation, screen captures are enough to provide evidence that the software works as intended.


Benefits of CSA over CSV

Moving from traditional CSV to CSA represents a cultural paradigm shift concerning validation, slashing the documentation, project timelines, and costs for life science companies implementing software.

In a December 2020 Medical Device Innovation Consortium (MDIC) Case for Quality Forum, the FDA and industry partners highlighted the results companies have achieved with this approach, including:

  • 90% or more reduction in test script and tester errors

  • Validation time cut 50% or more, with faster implementation as a whole

  • Lower overall project cost

  • Higher morale, quality, and productivity

  • More time for critical thinking as opposed to generating documentation

QMS Validation Services: What to Look For

It’s important to note that the right level of documentation is still necessary to avoid regulatory scrutiny. Follow the FDA audit perspective of, “If it’s not documented, it didn’t happen.”

In terms of QMS software, companies should look for the supplier’s out-of-the-box (OOB) technical and documentation that includes:

  • User/functional requirement specification

  • Risk assessment

  • Functional design specification

  • Requirement traceability

  • Reusable operational qualification (OQ)

  • Reusable performance qualification (PQ)

  • Validation summary report

The QMS should also come with technical and validation document templates that can be customized for your configured QMS, including templates for:

  • User/functional requirement specification

  • Risk assessment

  • Functional design specification

  • Requirement traceability

  • Configuration document

  • Operational qualification (OQ)

  • Performance qualification (PQ)

Automated validation is also a must for simplifying test management within configured solutions, allowing companies to deploy solutions faster while keeping the QMS in a state of control.

Finally, it’s important to ask questions about how professional services and validation services teams work together. Avoiding delays and costs requires a collaborative handoff between the two, with validation services part of the process from the beginning (even during project quoting).

The transition from computer systems validation to computer systems assurance is underway, and companies are already starting to see the benefits in terms of time and cost during software implementations. By enabling a risk-based approach using critical thinking, the FDA hopes to grow the adoption of technology that benefits patients, including QMS, product lifecycle management (PLM), laboratory information management system (LIMS) software, and more.

CSA does not replace CSV, nor does it contradict it. Instead, they need to coexist and be used appropriately. CSA is a simplified strategy for achieving CSV, allowing companies to spend more time testing systems based on risk than creating documentation—with better results to show for it.